Company record
Erbe USA Inc
0 FDA warning letters since 2023 and 4 recalls in the last 12 months.
Get an email the day FDA posts anything new about Erbe USA Inc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Aug 19 | Class II | Erbe USA Inc Device |
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000 A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture. |
| Apr 1 | Class I | Erbe USA Inc Device |
Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use Probes may rupture/burst during activation |
| Apr 1 | Class I | Erbe USA Inc Device |
Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use Probes may rupture/burst during activation |
| Apr 1 | Class I | Erbe USA Inc Device |
Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use Probes may rupture/burst during activation |